Hioki-LR8511 Wireless Universal module
Hioki-LR8511  Wireless Universal module

Hioki-LR8511 Wireless Universal module

Leverandør: Hioki E.E. Corporation
Varenummer: Hioki-LR8511
PDF Download: Hent brochuren her
Total er excl. moms: DKK 11.830,00
Antal:
tam_new_but2

WIRELESS UNIVERSAL UNIT LR8511

Logging Multi-point Data Has Never Been So Easy with a Wireless Logger

The Hioki LR8510/LR8511 are input modules of a wireless multi-channel data logger of voltage, temperature, resistance and humidity using Bluetooth® technology to send measured data to a central logging station. LR8410 series Wireless logger.

The LR8510 alone is not capable of making measurements. Wireless logging station Model LR8410-20 is necessary to measure. The main unit and input modules are not bundled with the Battery Pack Z1007. Thermocouples are not provided by HIOKI, and must be purchased from a separate vendor.
Countries and regions where wireless operation is currently supported: Japan, U.S.A., Canada, EU, Norway, Switzerland, Turkey, Russia, Vietnam, and India.
These products emit radio waves. Use of radio waves is subject to licensing requirements in certain countries. Use in countries or regions other than those listed above may constitute a violation of law, exposing the operator to legal penalties.
Bluetooth® is a trademark of Bluetooth SIG, Inc. and licensed for use by HIOKI E.E. CORPORATION.

 

• This is a input module to capture logging data using Bluetooth® wireless technology. Install logging modules in hard-to-reach locations (over line-of-sight distances of up to 30 meters *1 )
(*1) The presence of obstructions may shorten this range
• Choose an input module based on the parameters you wish to measure (15-channel type)
• Easily add up to 7 input modules wirelessly to keep your environment free of tangled wires (for a total of up to 7 modules for the LR8410 main unit, max. 105 channels when using 15-channel modules)
• 100 msec simultaneous sampling across all channels using rapid scanning method
 

■ Basic specifications (Accuracy guaranteed for 1 year)

Supported model Model LR8410-20
Control and communications Between instrument and unit: Bluetooth® 2.1 + EDR, Communications range: 30 m (98.43 ft), Security: SSP
Measurement parameters [No. of channels] 15 analog channels; isolated scanning method input (4 terminals: push-button type)
[Voltage] ±10 mV to ±100 V, 1-5 V f.s., max. 500 nV resolution
[Temperature: Thermocouples] -200 ˚C to 2000 ˚C (depends on sensor), Thermocouples (K, J, T, or other), max. 0.01 ˚C resolution
[Temperature: Pt 100, JPt 100 sensor] -200 ˚C to 800 ˚C, max. 0.01 ˚C resolution (not isolated between channels)
[Resistance] 0 Ω to 200 Ω f.s., max. 0.5 mΩ resolution (not isolated between channels)
[Humidity] 5.0 to 95.0 % rh (use with optional sensor), 0.1 % rh resolution (not isolated between channels)
[Max. rated voltage between isolated input channels] 300 V DC
[Max. allowable input] ±100 V DC
[Max. rated voltage from isolated terminals to ground] 300 V AC, DC
Operating temperature -20 °C to +60 °C (-4 °F to 140 °F)
Power supply [AC adapter] Using the AC adapter Z1008 (100 to 240 V AC, 50/60 Hz), 23 VA Max. (including AC adapter), 7 VA Max. (exclusive of AC adapter)
[Internal battery] Using the Battery Pack Z1007 (optional accessory), 24 hours of continuous use (at 100 ms recording interval, 23 ˚C reference data), 120 hours of continuous use (at 1 minute recording interval, 23 ˚C reference data), 0.6 VA Max.
[External power] 10 to 28 V DC, 7 VA Max.
Dimensions and mass 150 mm (5.91 in) W × 90 mm (3.54 in) H × 56 mm (2.20 in) D, 320 g (11.3 oz)
Accessories Instruction manual ×1, AC adapter Z1008 (with supplied power cord) ×1, Bracket ×1, M3×4 mm screw ×2