Hioki-3274 Clamp on probe, DC-10MHz, 150A Needs
Hioki-3274 Clamp on probe, DC-10MHz, 150A  Needs

Hioki-3274 Clamp on probe, DC-10MHz, 150A Needs

Leverandør: Hioki E.E. Corporation
Varenummer: Hioki-3274
PDF Download: Hent brochuren her
Total er excl. moms: DKK 22.110,00
Antal:
tam_new_but2
Wide-band, large current probe allows direct input to oscilloscope

Needs a Power Supply (3272 or 3269)

 
  • Even with the wide DC to 10 MHz frequency band, current waveforms up to 150 A can be directly observed in applications such as power device and power-on in-rush current monitoring
  • Connects directly to oscilloscope BNC input terminal
  • Connects directly to the MEMORY HiCORDER to provide waveform monitoring
  • An optional power supply is available for use with oscilloscopes that do not provide probe power
 
Note: Connect and power up Clamp-on sensor to use in combination with the oscilloscope, or the MEMORY HiCORDER
 
 
 
Basic specifications (accuracy guaranteed for 1 year)
 
Frequency band width DC to 10 MHz (-3 dB)
Rise time =
Noise level 25 mA rms max. (band width limited to 20 MHz)
Continuous allowable input 150 A rms (requires derating at frequency)
Max. allowable peak input 300 A peak (non continuous), 500 A peak (pulse width: 30 μs or shorter)
Amplitude accuracy ±1.0 % rdg. ±1 mV f.s. (DC, 45 to 66 Hz, 0 to 150 A rms)
±2.0 % rdg. (DC, 45 to 66 Hz, 150 A to 300 A peak)
Output rate 0.01 V/A (use with a device having a 1 M-ohm input resistance or higher)
Max. rated voltage to earth 600 V CAT II, 300 V CAT III (insulated wire)
Core jaw dia. Ø20 mm(0.79 in)
Power supply ±12 V ±0.5 V, 5.5 VA max. (at Max. current measurement)
Dimensions, mass 176 mm (6.93 in)W × 69 mm (2.72 in)H × 27 mm (1.06 in)D, 500 g (17.6 oz), Sensor cable BNC terminal: 2 m (6.56 ft), Power cable: 1 m (3.28 ft),
Power plug: FFA.0S.304.CLAC42Z / LEMO inc.
Supplied accessories Carrying case × 1